HELIUM GAS EQUIPMENT
HELIUM PURITY METERS
TCD
OPERATION:
A thermal conductivity detector (TCD) gas analyzer measures gas concentration by measuring the changing thermal gas conductivity with respect to changing impurity concentrations.
SPECIFICATIONS:
· Sensitivity: Equivalent to 2% air in helium
· Accuracy: 0.8% full scale air in He
· Response time <30 s for most applications
· Operating Temperature Range: ambient (without temp. control option) ambient - 100ºC
FUEL CELL
FEATURES:
·Helium Purity Meter - remote controlled valve option.
·Provides a remote controlled (motorized) valve to switch between the sample gas and reference gas.
·Useful for automating the readings when the high purity requires that comparison with a reference gas is required.
·Helium Purity Meter - gas chromatography separation column option.
·Allows measurement of the relative concentration of oxygen and nitrogen in the contaminants.
Click for more info
·Measures oxygen, which typically relates to contamination of helium by air.
SPECIFICATIONS:
· Concentration: 0-100% Helium in air
· Pressure Range: Atmospheric (1 bar) to 1.4 bar
· Accuracy: ±1% Full Scale
· Response Time: 90% in less than 15 seconds
· Operating Temperature: 5-25C (59-104 degF)
· Minimum/Maximum Storage Temperature: -15-50C (5-122 degF )
· Optimal Storage Temperature: 41-77 degF (5-25C)
· Humidity: 0-95% non-condensing
·Life span: 9-12 months in open airPlease contact Quantum Technology Corp. for more information.